Check whether a replacement pH electrode fits the existing holder. Match the shoulder datum, sealing point and installed sensing-tip position on the drawing.
A matching pH range does not establish that an electrode will reach the sample or connect to the transmitter. Record the installed holder, measurement datum and electrical interface before requesting a spare.
Locate the reference surface
Measure insertion length from the reference surface on the holder drawing. State whether the assembly is fixed or retractable. Also identify the connector, cable and measurement technology. Two probes that look alike may require different transmitter inputs.
When the tip stops short
Suppose two drawings list the same length, one from the shoulder and the other from the cable end. Overlay their reference surfaces before comparing the tip position or deciding that the probes are interchangeable.
Enquiry checklist
| Item | What to record |
|---|---|
| Holder | Type and order code. |
| Geometry | Insertion datum, length and connection. |
| Signal | Electrode technology, cable and transmitter. |
Check the quotation Identify the part that remains installed after this purchase. Ask for the offered sensor, module or service kit to be paired with that exact installed model in the quotation.
Specify the electrode and seals included in the supply. A quotation for a complete retractable assembly has a different scope from a spare sensor.
Sample enquiry to ISATIS
Please quote the electrode for the attached holder and transmitter. Show the insertion reference and identify seals, cable and any adapter separately.