Identify the measuring principle, cell constant, temperature element, input module and holder before requesting a conductivity probe for a liquid analyzer.
The model on a liquid analyzer display identifies only part of the measuring chain. When a buyer asks for a conductivity probe for an FLXA21, the installed input, process sensor, holder and cable still need to be identified. Otherwise, a sensor quotation can easily be compared with an analyzer package.
Record the measurement principle and cell constant
Check the installation documents for contacting or inductive measurement. For a contacting cell, record the electrode arrangement and cell constant with its unit. A cell constant expressed in reciprocal centimetres is a different quantity from the process conductivity expressed in microsiemens per centimetre.
Identify the installed input
The FLXA21 uses different measurement configurations. Read the order code or installed module documentation, then record the temperature element and cable specification. A matching connector or a similar number of conductors is insufficient evidence of compatibility.
| Check | What to record |
|---|---|
| Sensor record | Electrodes, cell constant and temperature element |
| Transmitter record | Installed measuring input and module code |
| Mechanical record | Holder and insertion dimensions |
Check the quotation Identify the part that remains installed after this purchase. Ask for the offered sensor, module or service kit to be paired with that exact installed model in the quotation.
Review a proposed equivalent
Consider an installed two-electrode sensor and a quotation that says only “for FLXA21”. Ask for the exact sensor code, cell constant, temperature element, cable length and mounting arrangement. If the holder stays, compare its drawing with the proposed sensor diameter and insertion geometry.
Put the supporting records together
Send the probe identification, the transmitter input details, expected conductivity range and holder drawing. Include process and cleaning conditions separately where they differ. Have the quotation identify whether the cable, adapter and holder are included so the purchasing comparison uses the same scope.
Related equipment and quotations
- SC4A-S-AD-09-002-05-T1/PS/FF — 19 mm 2-electrode conductivity sensor
- FLXA21 — 2-wire liquid analyzer / transmitter