Match each pressure gauge serial number to its calibration certificate. A three-gauge example shows how to catch a duplicate record even when document counts match.
A certificate needs to be traceable to the instrument it describes. Range and model can be identical across several gauges, so match the actual serial number on each instrument to its own record.
Check each gauge as an individual item
Record model, range, units, serial, certificate identifier and relevant calibration points. Check the required dates and result format against the purchase order. Use the instrument marking rather than relying solely on a carton label.
Three gauges rated 0–10 bar
A delivery may contain three matching gauges and three certificates. A complete set of papers by count is still insufficient if one serial is repeated and another is missing. Reconcile each unit before closing the receipt.
Enquiry checklist
| Item | What to record |
|---|---|
| Instrument | Serial from the actual gauge. |
| Certificate | Certificate number and matching serial. |
| Coverage | Range, points and required results. |
Check the quotation Specify the evidence you need at acceptance. Link each record to the equipment tag or serial and ask which results will appear in the delivered report.
Keep the mapping in the receiving record and ask for any mismatched document to be corrected before the record is accepted.
Sample enquiry to ISATIS
Please provide a certificate and serial-number list for each of the three gauges. Identify the corresponding instrument, range and calibration points on every record.