Compare the pyrometer’s spot size at the working distance with the hot target. Check surface emissivity and any viewing window when specifying the measurement.
An infrared thermometer collects radiation from an area, and the reported temperature also depends on surface properties. Record target size and surface condition as well as the temperature range.
Check the spot at the installed distance
Use the offered model's optical diagram to compare measurement spot size with the actual target. Record the expected emissivity conditions, view angle and any window between the instrument and target.
A small hot part against a cooler background
If the target is smaller than the spot, the background contributes to the reading. A surface finish change or unsuitable infrared window can also affect the result even when the instrument range is unchanged.
Enquiry checklist
| Item | What to record |
|---|---|
| Target | Size, material and surface finish. |
| Optics | Distance, spot size and view angle. |
| Window | Material and transmission for the proposed spectral band. |
Check the quotation Mark the viewing position, target and intervening window, if present, on one sketch. Ask which optical components and calibration assumptions the offered configuration includes.
Have the proposal state the assumptions used for emissivity and optical access. Confirm the arrangement on a representative target before accepting the selection.
Sample enquiry to ISATIS
Please quote a pyrometer for the attached target and viewing geometry. State spot size at the working distance and the assumed surface and window properties.