Will both the reference probe and the sensor under test fit the dry-block insert? Compare bore diameters, usable depth and probe length before placing the order.
A dry-block insert must suit the probe’s sheath diameter and the length that can actually enter the block. “For a Pt100” identifies the sensing principle, not the required bore.
Read the bore drawing against the probes
Measure each sheath and its straight insertion length to the first mechanical obstruction. A comparison also needs a suitable position for the reference probe. Excessive clearance or inadequate immersion can degrade the comparison. Use the recommendations for the actual block and probe dimensions.
Two probes on one calibration bench
A 6 mm process probe and a reference of a different diameter may both need to be immersed at the same time. Several identical holes are not necessarily helpful. Specify the required bore pattern and identify which position is for the reference.
Enquiry checklist
| Item | What to record |
|---|---|
| Working probe | Sheath diameter and usable straight length. |
| Reference | Diameter and required immersion for the comparison probe. |
| Insert | Part number, bore pattern and operating range. |
Check the quotation Draw the temperature loop from sensing element to receiving input. Circle the part being purchased and mark the items that will stay in service, including any existing cable.
At receipt, compare the insert code and drawing with the quotation. The block’s maximum temperature does not establish performance with a short probe.
Sample enquiry to ISATIS
Please quote an insert for our existing dry block and the two probes listed. Supply the insert code, bore diameters and depths, and the applicable temperature range.