Replacing remote Coriolis electronics? Identify the installed sensor, transmitter and cable, then compare an electronics-only quote with a complete meter package.
The name on the display is not necessarily the model of the sensor in the pipe. Identify which components will remain installed before requesting a Coriolis spare.
Collect both nameplates
Record the sensor code, process connection size and wetted materials. For the transmitter, record model, supply and outputs in use. On a remote installation, identify the interconnecting cable, its length and any intermediate electronics. These define what needs a compatibility check.
Two offers with different supply boundaries
If the workshop needs only the electronics and display, a complete meter price is not directly comparable. Conversely, a “complete transmitter” may omit both sensor and cable. Ask for a parts list against each option.
Enquiry checklist
| Item | What to record |
|---|---|
| In the line | Sensor nameplate and process connection. |
| At the wall | Transmitter nameplate and cable interface. |
| At delivery | New items, retained items and compatibility evidence. |
Check the quotation Send the line tag and mark exactly which item is being replaced on the existing meter drawing. Ask the bidder to identify any change to installation dimensions or to the signal used by the control system.
Request confirmation of compatibility and any configuration or commissioning needed after replacement. Serial numbers from both ends reduce the risk of reviewing the wrong generation.
Sample enquiry to ISATIS
Please quote replacement electronics for the sensor shown on the attached plate. Confirm the existing cable can remain and identify any interface component or configuration work required.