Specify calibration points in engineering units. The midpoint of 4–16 bar differs from the midpoint of 0–16 bar, even though both are labelled 50%.
A calibration service request should give actual point values and units as well as any percentages. The same percentage can describe different physical pressures when the lower range value changes.
Calculate from both ends of the span
For a specified fraction of span measured from the lower limit, add that fraction of the difference between upper and lower limits to the lower limit. Give the required ascending or descending sequence and result format.
Two ranges at their midpoint
Halfway through 0–16 bar is 8 bar. Halfway through 4–16 bar is 10 bar. A request listing only 50% could therefore lead to a valid point for the wrong configured range.
Enquiry checklist
| Item | What to record |
|---|---|
| Range | Configured lower and upper values. |
| Points | Actual values with units and sequence. |
| Record | Required results, uncertainty and instrument identifier. |
Check the quotation Give one actual test point and request the applicable limit in pressure units at that point. List the pump, reference and connection kit separately if the quotation covers more than one item.
Ask for confirmation of the point list before the instrument is dispatched. Identify any difference between the nameplate range and configured working range.
Sample enquiry to ISATIS
Please calibrate the instrument at the attached pressure values in the indicated order. Record the configured span, units and results at each point.