Which pressure does a dew-point reading refer to? Track the sample from the line through pressure reduction and specify the basis required at the sensor.
Gas taken from a pressurised line may reach the dew-point sensor after pressure reduction. A dew-point figure without its pressure reference can lead to a misleading comparison between the process target and the instrument reading.
Mark the cell on the sampling sketch
Show the take-off, reducer and measuring cell, with pressure at each relevant point. Identify the moisture quantity and pressure basis required for reporting, together with the normal sample flow and temperature.
One gas line, two pressure references
An instrument measuring downstream of a reducer does not necessarily report the same dew-point temperature as a sensor at line pressure. Record the actual measurement pressure even when the purchase specification cites main-line pressure.
Enquiry checklist
| Item | What to record |
|---|---|
| Process | Line pressure and target moisture basis. |
| Cell | Pressure at measurement and expected range. |
| Sampling | Reducer, sample flow and relevant temperatures. |
Check the quotation Send a sketch showing the sample take-off, conditioning equipment and measurement point. Mark the items included in this enquiry so that a price for one component cannot be mistaken for a complete sampling system.
Ask whether a proposed reported value is measured directly or calculated from other inputs. Have the pressure reference written into the acceptance criteria.
Sample enquiry to ISATIS
Please quote the dew-point measurement package for the marked sampling point. Give the cell pressure basis and explain how the requested reported value is obtained.