In natural gas sales you sell energy, not volume. The gas chromatograph is the instrument that measures that energy.

What natural gas contracts ultimately trade is energy, not just cubic metres. Two gases with identical volume can have very different heating values, and on a large custody transfer station a small error in heating value is worth more each year than the analyser that measures it. The instrument that determines gas composition, and therefore heating value, is the process gas chromatograph (GC).

This guide covers how an online GC works, what it measures, how heating value is calculated, and the practical factors that decide whether its results can be trusted in a commercial dispute.

How a chromatograph works

At a fixed interval the GC’s sample valve injects a small, precisely measured volume of gas into one or more separation columns. A carrier gas, usually helium, pushes the sample through the columns, and each component leaves at a different time depending on its affinity for the stationary phase. A thermal conductivity detector (TCD) records each component as it elutes; the area of each peak is proportional to its concentration.

Natural gas GCs normally use several columns and valve switching. Heavier hydrocarbons are often backflushed as a single C6+ group, which shortens the cycle. A typical C6+ analysis completes in a few minutes, so the flow computer receives a fresh composition several times an hour.

What is measured?

Analysis Components Typical use
C6+ N2, CO2, C1 to C5 individually, C6+ as a group Pipeline-quality gas, custody transfer
C9+ extended As above, with C6, C7, C8 and C9+ separated Rich gas, hydrocarbon dew point estimation
Sulfur species H2S, mercaptans, COS Gas quality specifications, sour gas
Special H2, O2, He Hydrogen blends, biomethane, LNG

Sulfur components at ppm level usually need a different detector, such as a flame photometric or sulfur chemiluminescence detector, or a dedicated H2S analyser alongside the main GC.

From composition to heating value

With the molar composition, the flow computer or the GC itself calculates the following according to ISO 6976 or GPA 2172:

  • Gross and net heating value (GHV / NHV)
  • Relative density and density at standard conditions
  • Wobbe index, for burner interchangeability
  • Compressibility factor (Z) for AGA 8 or GERG-2008 calculations

Two details cause recurring disputes. First, the reference conditions for combustion and metering, such as 15 °C or 60 °F, must match the contract. Second, the contract must state which edition of the standard applies, because ISO 6976:2016 changed some physical property values compared with the 1995 edition.

What affects accuracy

  1. Sampling system. Most GC errors start before the gas reaches the analyser. The sample must be representative, taken from a properly located probe, and kept above its hydrocarbon dew point with heated lines and a heated pressure regulator. ISO 10715 and API MPMS Chapter 14.1 give the requirements.
  2. Lag time. A long, oversized sample line delays the analysis. A fast loop with a bypass to a low-pressure point keeps transport time short.
  3. Calibration gas. A certified working reference gas close in composition to the process gas, kept above its own dew point and within its validity date.
  4. Oven temperature and carrier gas. Stable oven temperature and carrier pressure keep retention times and response factors repeatable.
  5. Validation. Periodic checks with an independent reference gas, and a performance evaluation according to ISO 10723, show whether the analyser is still within its stated uncertainty.

Calibration and validation in practice

An online GC normally calibrates automatically, for example once a day, by analysing the working reference gas and updating response factors. Good practice is to alarm and reject a calibration when response factors move more than an agreed limit, instead of silently accepting it. Trend the response factors, retention times and unnormalised total over months; a drifting total is often the first sign of a sampling or valve problem.

Applications

Custody transfer metering stations, gas plant inlets and outlets, power plants, compressor stations, gas blending and LNG quality control are the main applications of process GCs. In each case the GC is part of a system: probe, sample conditioning, analyser shelter, utilities, calibration gas and the flow computer that uses the results.

Frequently asked questions

Can a GC measure water or hydrocarbon dew point directly?

Water is usually measured with a separate moisture analyser. Hydrocarbon dew point can be estimated from an extended C9+ analysis with an equation of state, or measured directly with a dedicated dew point analyser where the contract requires it.

Why is the unnormalised total important?

The GC reports a total close to 100% before normalisation. A total that drifts away from 100% points to sample pressure, valve or calibration problems that normalisation would otherwise hide.